Issue 1
2-2 Prepare the instrument K0460 - [EN] English
2.5 The battery
WARNING
• This instrument uses the GE specified, Ni-MH battery
pack. To prevent an explosion or fire, do not short
circuit, do not disassemble, keep it safe from damage.
For operating conditions, see Table 11-1.
• Charge the GE specified battery in a safe area, using the
GE supplied charger and cradle.
• To prevent battery leakage or heat generation, only use
the battery charger and cradle in the temperature range
0 to 40°C (32 to 104°F). For operating conditions, see
Table 11-1.
For a full battery specification, refer to Table 11-1.
2.5.1 Battery condition
On receipt of a new DPI 620-IS calibrator, the battery will be
partly-charged. We recommend that it is fully-charged
(Section 2.5.3).
Charge indications After switching on, the battery symbol at the top of the display
shows the charge condition in 10% increments.
To get an accurate indication (1% increments), use the
Configuration menu; see Section 2.8.2 (Procedure to see the
instrument status).
2.5.2 Install the battery
Step Procedure
1. When the power is off, loosen the five screws (a)
and remove the cover (b).
If necessary, turn the instrument over and collect
the discharged battery.
If the battery does not hold a charge, discard it
safely. Obey all the local health and safety
procedures.
2. Install the new fully-charged battery in the
compartment.
3. Refit the cover.