[EN]  English - K0460 Datalog operation   6-1
Chapter 6: Datalog operation
6.1  Introduction
This section gives examples of how to log measurements with 
time and date over a set time period or on a key press.  Logged 
data is stored in a user defined file.  The instrument logs all the 
tasks currently enabled.
Stores this data internally but can be downloaded (transferred) 
through the mini Type B USB serial port.  The data transfer to 
the PC uses Active Sync.
To enter data logging select configuration menu and press data 
logging.
6.2  Set-up
To set the data logging process:
After data logging the data file can be used in the data logging 
menu as follows:
Selection Action
Filename Enter name using three-page screen alpha/numeric 
key-pad.
Trigger Set to Key Press to activate on pressing of bottom 
right key or Periodic set to activate on time set in 
period.
Period Sets the time period of data recording using numeric 
keys and HH:MM:SS format.
Start Starts data logging.