[EN] English - K0460 Documenting functions 7-1
Chapter 7: Documenting functions
7.1 Introduction
This section gives examples of the documenting functions
available with the DPI 620-IS calibrator. There are two options:
Analysis (Section 7.2): This function compares the data from
two Channels on the DPI 620-IS calibrator: the device under
test (DUT) and a reference instrument.
It calculates the % span or % reading error for the device and
then gives a pass or fail indication.
Run Procedure (Section 7.3): This uses the Intecal
Calibration software (Optional item) to download a calibration
procedure.
Before starting:
Read and understand the “Safety” section.
Do not use a damaged instrument.
Note: Use only original parts supplied by the manufacturer.
7.2 Analysis
To use the Analysis function, set the functions in the two
channels for comparison (Section 2.8.4). Then use these
procedures:
2
1
TAP
Two Tasks
1
2
3
4
Set one channel as the Input (the DUT) and the other as the Reference.
Set the options A to E.
5
Documenting
Input
Reference
7
6
A
B
C
D
E
8
Continued