Issue 1 
8-8   HART
®
 device operations K0460 - [EN]  English
  2. Set-up the HART
®
 device, if loop power required, use CH2 for 
current (24V).
      If the loop is already powered and there is an external HART
®
 
resistor, turn off CH2.  Connect CH2 mA+ and mA- connections 
to the HART
®
 resistor, see Section 1.3.
If required, turn on the internal HART
®
 resistor.  This can be 
selected by pressing   icon.
Press the HART
®
 button and the application starts.
3. Complete the electrical connections and continue with the 
HART
®
 menu operation; see Section 8.4.
4. Complete the electrical connections and continue with the 
HART
®
 menu operation; see Section 8.4.
8.3.2 External loop power
In this example, there is an external power supply. 
Set the functions for channel 2 and HART:    
5.   Set the HART
®
 Resistor. 
Measure current on 
channel 2 (CH2)
Range: ±55 mA
HART Function is set
250Ω resistor is set
e
1
a
b
d
c
Set HART
Section 2.8.4
Set Current; Measure
Section 2.8.1
Home display
T