3-4 Electrical operations K0460 - [EN] English
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
3. To continue, set the applicable output value; see
Section 2.9.2.
3.2.5 Example procedure: Measure or source frequency signals
These examples (A and B) show Channel 1 (CH1) set up to
measure or source a frequency. This includes Hz, kHz and
counts (cpm or cph).
The selection of units sets the available range, for example:
Example 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
Source DC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV (DC)
Automation: Nudge
(Section 2.9.2)
B 1
3
2
Hz = 0 to 1000 Hz kHz = 0 to 5 kHz
21
Measure frequency on
channel 1 (CH1)
Range: 0 to 5 kHz
Trigger level: 2.5 V
A
3
Source frequency on
channel 1 (CH1)
Range: 0 to 5 kHz
Waveform: Triangle
Amplitude: 5.0 V
Automation: Nudge
(Section 2.9.2)
2B
4
1
3