[EN] English - K0460 Electrical operations 3-3
Issue 1
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure or source operation.
3. Source only (Automation): Set the applicable output value;
see Section 2.9.2.
3.2.3 Example procedure: Measure DC voltage
These examples (A and B) show Channel 1 (CH1) set-up to
measure a DC voltage.
Note: Using Channel 2 (CH2) connectors, use Channel 2 to
measure these ranges.
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure operation.
3.2.4 Example procedure: Source DC voltage (CH1)
These examples (A and B) show Channel 1 (CH1) set-up to
source a DC voltage.
2
Measure DC millivolts on
channel 1 (CH1)
Range: ±2000 mV (DC)
1
B
Measure DC volts on
channel 1 (CH1)
Range: ±30 V (DC)
A
1
21
3
Source DC volts on
channel 1 (CH1)
Range: 0 to 12 V (DC)
Automation: Nudge
(Section 2.9.2)
A