2-18 Prepare the instrument K0460 - [EN] English
2.10 The Advanced
menu options
The Advanced menu provides settings and use of options:
2.10.1 Advanced: Calibration options
There are two calibration menus that can be used:
Menu to calibrate the
DPI 620-IS
1. Enter the calibration PIN: 4321.
2. Tap the “Accept” button and continue with these operations:
Calibrate the different measure and source channels on the
DPI 620-IS calibrator.
Contact us at: www.gesensinginspection.com
Change the calibration PIN and then enable it, confirm the
new PIN.
Set the next calibration date and/or set the Notification
option. If Notification is On and calibration is overdue, the
display shows “Calibration due ... ” message at the end of
the power on sequence.
Settings Calibration
Files Options
2
1
See Section 2.10.1.
Hart On or Off; see Chapter 8. Standards sets
IPTS 68 or ITS 90; see Chapter 3
Lists options installed.
See Section 2.10.2.