Issue 1
7-2 Documenting functions K0460 - [EN] English
Input and Reference
options
Channel type : Input or Reference
Scaling (Input only): Values for Reference High and Low
and Input High and Low. This sets the scale for the Analysis
function.
Error Type (Input only): % Span or % Rdg (Reading)
Linearity (Input only): Linear or Square Root
Tolerance (Input only): an applicable value for the
pass/fail calculation
When the Input and Reference options are set, start the
analysis.
7.3 Run a procedure
The Intecal calibration Procedure contains all the values to
calibrate a device under test (test points, ramp time). Use the
same calibration procedure for all the applicable devices under
test.
To use the Run Procedure function, the following items are
required:
a copy of the Intecal Calibration Software. This is available
on the website at: www.gesensinginspection.com
There is a 30 day free trial.
IO620-USB-RS232 (Intecal version 5 only). An adaptor to
connect the DPI 620-IS calibrator to the RS232 interface of
a computer.
a DPI 620-IS calibrator device driver: available as a
download from our website.
A
B
C
D
E
Set applicable values. When the Analysis
work is complete, tap on exit.
Start the
Analysis
9
10
Continued
from step 8
in specified
tolerance limit
not in specified
tolerance limit