[EN] English - K0460 Calibration procedures 10-11
Issue 1
10.11 Procedures
(CH1): Resistance
(source)
1. Connect the applicable calibration equipment in Table 10-1;
for example: see Section 3.2.9 (Example procedure:
Measure or simulate an RTD (or Resistance)), example B.
2. Let the equipment get to a stable temperature (minimum: 5
minutes since the last power on).
3. Use the calibration menu (Section 10.3) to do a two-point
calibration for each range.
• Range: 0-400Ω
• Range: 400Ω-2000Ω
• Range: 2kΩ-4kΩ
4. To make sure the calibration is correct, select the applicable
Resistance (source) function; see Section 3.2.9, (B).
5. Set the specified values (Table 10-11). Make sure the error
is in the specified limits.
Table 10-11: Resistance (source) error limits
Ohms
(Ω)
Excitation
(mA)
Calibrator
uncertainty (Ω)
Permitted
DPI 620-IS error
(Ω)
0 0.1 0.0014 0.014
100 0.1 0.0016 0.038
200 0.1 0.0021 0.062
300 0.1 0.0028 0.11
400 0.1 0.0035 0.31
1000 0.1 0.008 0.55
2000 0.1 0.016 0.86
4000 0.1 0.032 1.1