[EN] English - K0460 Pressure indicator operation (MC 620-IS) 4-3
Issue 1
4.4 Measure
pressure
When the pressure indicator assembly is complete
(Section 4.2.1), use the menus to set-up the necessary
operations.
4.4.1 Procedure overview
To use the pressure indicator, complete these procedures:
• Set the calibrator functions to use on the display; see
Section 2.8.3. This includes:
a. Pressure function (P1 and/or P2): item .
b. CH1: Channel 1 electrical function (measure or source).
c. CH2: Channel 2 electrical function (measure or source).
d. Other functions (maximum: 6 functions).
• If necessary, change the Units for the function: item ;
see Section 2.8.5.
• If necessary, set a Utility for the function: item ; see
Section 2.8.6.
a. Max/Min/Avg
b. Switch Test
c. Leak Test; see Section 4.4.2.
• If necessary, change the Settings for the pressure function:
item .
a. Process (Tare, Alarm, Filter, Flow, Scaling);
see Section 2.9.1.
b. Leak Test (Only when the Utility is set); see
Section 4.4.2.
c. Zero Recommend - zero a gauge sensor before use; see
Section 4.4.3.
d. Display If necessary, change the number of digits in the
display reading.
Step Procedure
4. Re-attach the adaptor to the MC 620-IS carrier and
tighten it until it is hand tight only.