[EN] English - K0460 Electrical operations 3-7
Issue 1
3. If necessary change the Source Settings and continue with
source operation.
Automation: Set the applicable output value; see
Section 2.9.2.
Values to set:
RTD Type (Set the applicable RTD); see Tabl e A1 (front
cover) for the available options.
3.2.8 Example procedure: Measure or simulate a thermocouple (or TC mV)
These examples (A and B) show Channel 1 (CH1) set-up to
measure or simulate a thermocouple temperature.
To measure or simulate TC millivolts, set the TC mV function.
Example A 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
21
Measure a TC temperature
on channel 1 (CH1)
Range: 1372 °C
TC Type: K Type
Cold Junction: 27.43
A
3
Simulate a TC temperature
on channel 1 (CH1)
Range: 1372 °C
TC Type: K Type
Cold Junction: 38.2
Automation: Nudge
(Section 2.9.2)
2B
4
1
3