[EN] English - K0460 Electrical operations 3-5
Issue 1
3. If necessary, change the Trigger Level (Settings) and
continue with the measure operation.
Values to set:
Mode (Automatic/Manual);
Manual Level (trigger level value)
Example 1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections.
3. If necessary change the Source Settings and continue with
source operation.
Values to set:
Waveform (Square, Triangle, Sine);
Amplitude (Amplitude value)
4. Automation: Set the applicable output value; see
Section 2.9.2.
3.2.6 Example procedure: Measure or source current (24V loop power)
These examples (A and B) show Channel 2 (CH2) set up to
measure or source a current with 24V loop power.
1. Set the applicable software options; see Section 3.2.1
(Procedure overview).
2. Complete the electrical connections and continue with the
measure or source operation.
3. Source only (Automation): Set the applicable output value;
see Section 2.11.2.
3
B
21
Measure current on
channel 2 (CH2) with
24V loop power
Range: ±55 mA
A
Maximum: 24 mA
Source current on
channel 2 (CH2) with
24V loop power
Range: 0 to 24 mA
Automation: Nudge
(Section 2.11.2)
2B
3
1