Issue 1 
8-6   HART
®
 device operations K0460 - [EN]  English
mA
A
n
a
l
o
g
 
t
r
i
m
 
• Type the Analogue value of Primary Variable mA 
current (actual calibrated value) into the box 
using the key-pad and then press Set.  Repeat 
this process with 20 mA selected.  This will 
accurately calibrate the transmitter’s output 4 
to 20 mA loop current. 
Sensor Trim    
• Using the DPI 620-IS, a sensor trim on most Primary Variables 
can be performed without having to connect any external 
reference meters.  If the Primary Variable is a pressure device 
use the DPI 620-IS and PV 62x-IS pressure generation stations 
to provide the pressure source.  Alternatively, use DPI 620-IS 
and MC 620-IS to accurately measure pressure.
• Navigate to the calibration folder and select the sensor trim 
method.  Follow the on-screen instructions.
 
• In this example a lower trim has been selected.  
The DPI 620-IS is set-up to source a Primary 
Variable to the transmitter.  CH1 has been 
configured to output 0°C.  This can be done by 
selecting the Switch App button at the bottom 
of the method window and configuring the DPI 
620-IS to simulate an RTD at 0°C.   The Analogue 
Value of the Primary Variable can be seen by 
selecting CH1 on the drop-down menu.
• Type the Analogue Value of the Primary Variable 
(actual calibrated value) into the box using the 
key-pad and then press Set.  Repeat this 
process with an upper trim value.  This will 
accurately calibrate the transmitter’s sensor 
output.