[EN] English - K0460 Calibration procedures 10-15
Issue 1
5. Make sure the error is in the specified limits; see the tables
for Gauge Ranges or Absolute Ranges given in the
datasheet (supplied on the CD: P/N UD-0002); use the
values in column “Total uncertainty ... “.
The specified values include an allowance for temperature
changes, reading stability for one year, and the uncertainty of
the standard used for calibration.