8 - 40 Section 8-4 - Keyboard Block
8-4-12-4 Removal Procedure (cont’d)
3.) Disconnect two cable connector. Unscrew ten(10) screws (1-10) to remove the Menu Encoder
Assy. Refer to Figure 8-42.
4.) Perform the following functional tests. If all are successful, include the debrief script provided below.
8-4-12-5 Mounting Procedure
Install the new parts in the reverse order of removal.
Figure 8-42 Unscrew 4 screws from Menu Encoder Circuit Board
Table 8-19 Functional Tests
Service Manual
Section
Functional Test / Diagnostic Test Debrief Script
Section 4-3-1
Power On/Boot Up
“Service Manual, Direction
2380207, Rev 7+, Section 8-4-12.
Equipment passed all required tests and is
ready for use. “