DIRECTION 2380207, REVISION 7 LOGIQ™ 5 PRO SERVICE MANUAL
8 - 154 Section 8-9 - Power Block
8-9-1-4 Removal Procedure (cont’d)
3.) Unscrew ten (10) screws (11-20) to remove AC Power assy. Refer to Figure 8-173.
4.) Remove the AC Power Assy.
Figure 8-173 Removing the AC Power Assy
WARNINGWARNING
The weight of the AC Power Assy is approximately 20 kg. Two person is needed in the
next step.