Configuration
2026419-033E CIC Pro™ 6-35
6. Verify the Automatically adjust clock for daylight saving changes box is not
checked.
CAUTION
DATA CORRUPTION—Do not select the Automatically adjust
clock for daylight saving changes check box. You must manually
make Daylight Saving Time changes in the monitoring system. This
change may affect the CARESCAPE Network time and date
parameter of some patient data and FD data corruption.
7. Click Apply to apply your changes.
CAUTION
Do not make any changes on the Internet Time or Date & Time tab
to change the time. The time change request must be made only
from Webmin. Otherwise, the time change request will not be sent to
the Time Master.
8. Click OK to close the window.
9. Restart the CIC Pro center. From the Windows taskbar, click Start > Shut Down
> Restart and press Enter.
CAUTION
LOSS OF FULL DISCLOSURE DATA — Failure to reboot the
CIC Pro center after changing the time zone results in: 1) The loss of
stored full disclosure data. 2) The discontinuance of full disclosure
data collection. 3) The inability to access the full disclosure
function.
10. Run the Check Centrals utility to ensure that there are no time zone or network
errors. See Pre-configuration instructions on page 6-2 for instructions.