Configuration
2026419-033E CIC Pro™ 6-49
5. After making your selections, choose one of the following:
Click OK to apply your changes and close the CIC Setup window.
Click Apply to apply your changes without closing the CIC Setup window.
6. Verify the new settings prior to monitoring patients.
Set the Display Configuration (non-mirror CIC Pro centers)
The Display Configuration tab allows you to format the CIC Pro center’s multi-
patient viewer with the required number of patient windows.
The amount of displayed patient data is dependent on the total number of patient
windows displayed in the multi-patient viewer. It is also dependent on the Auto
Display Button settings on the Display Configuration tab (e.g., Maximizing the
Waveform Length or Maximizing the Number of Waveforms).
Modifications to Display Configuration are subject to licensing restrictions.
Licensing determines the number of beds displayed on the CIC Pro center screen.
Modifying the display to show more than the licensed number of beds will result in
blank slots in the display.
NOTE
You must remove admitted beds from the display before you can select a Display
Configuration that would eliminate those patient slots from the display. In the
multi-patient viewer, right-click on the patient bed you want to remove and
choose Select Care Unit then Bed Number > None.
Unit License Default: Full
Disclosure License Type
NOTE
If the default does not match the actual license, full disclosure does not work.
Display a list of the full disclosure license type.
Offline Storage
WARNING
POTENTIAL DATA LOSS—Do not allow a NO COMM (patient offline) event to exceed the time limit selected
in the Offline Storage setting. The patient’s full disclosure data is deleted if the time limit is exceeded.
Set the amount of time stored in Offline Storage. This exists as a precautionary measure when contact with
a monitor has been lost. This feature applies most often when the monitor is being used in the Rover
application, since the monitor moves (roves) from room to room, on and off the network.
If a discharge or admit cycle occurs off the network for a bed with full disclosure data collection, the full
disclosure sub-system is unaware that the patient has changed. When the bed returns to the network, full
disclosure will present the data from the discharged and admitted patients as part of the more recent patient’s
data. Offline Storage helps to prevent this by assuming that the patient has been discharged if the monitor is
off the network for greater than the number of minutes selected with the Offline Storage control.
To modify the Offline Storage period, choose the time period from the drop-down list and select Yes to
confirm that you want to change the time period.
Start Data Storage Designate how full disclosure is enabled for patients at the time of admission.
Bed List Lists beds for which full disclosure data is automatically stored.
Restore button Clear any changes you made to the Full Disclosure Defaults settings and revert to the previous settings.
Option Function

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MP100 Series and is the answer not in the manual?

GE MP100 Series Specifications

General IconGeneral
BrandGE
ModelMP100 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals