12-4 CIC Pro™ 2026419-033E
Upgrade software
Prepare the service PC
This section describes the procedures required to prepare the service PC for installing
software packages to networked CIC Pro centers:
Connect the service PC to the CARESCAPE Network IX network on page 12-4.
Set up the service PC’s network properties on page 4-8.
Start the software transfer utility on page 12-4.
Enter the CARESCAPE Network IX network addresses of CIC Pro centers to be
updated on page 12-5.
Connect the service PC to the CARESCAPE Network IX network
WARNING
LOSS OF ALARMS — When using a service PC to install or reload
software onto the CIC Pro center, do not connect the Ethernet cable
from the service PC to the CIC Pro centers CARESCAPE Network
MC network port. You must use the CARESCAPE Network IX
network to install or reload software.
Connect an Ethernet cable from the service PC’s Ethernet port to the facility’s
CARESCAPE Network IX network.
Start the software transfer utility
1. Insert the DVD containing the GE Software Distribution Utility into the service
PC’s DVD drive.