Configuration
2026419-033E CIC Pro™ 6-53
6. To change an existing hardwire bed, telemetry bed, or telemetry transmitter/
transceiver, complete the following tasks:
a. Use the scroll bar to locate the item you want to change.
b. Click on the item to display an editing window.
c. Press Enter to add the entry to the list.
7. After making your selections, complete one of the following tasks:
Click OK to apply your changes and close the CIC Setup window.
Click Apply to apply your changes without closing the CIC Setup window.
8. Verify the new settings prior to monitoring patients.
Set up locked beds
When logged into the Service mode, the CIC Pro center can be configured with the
bed names either in locked or unlocked mode. When locked, the bed name is
permanently allocated to a particular slot on the CIC Pro center and users are unable
to move the bed to another slot.
NOTE
It is possible to admit a patient to a window with a bed name that is locked to
NONE. To avoid duplication of patient waveforms, a window locked as NONE
should not be used to admit a patient.
Complete the following procedure to lock or unlock a bed:
1. From the multi-patient viewer, click Setup CIC.
2. Log on to Setup CIC with service access on page 4-5.
3. Position the mouse pointer in the patient’s waveform window and right-click.
4. Click to choose either LOCK or UNLOCK. The change takes effect immediately.
(Chinese only) Set the pressures unit-of-measure
NOTE
This option is only available on CIC Pro centers running software v5.0.6 or later
configured for the Chinese language. Pressure values may be displayed in kPa
when configured for the Chinese language.
Complete the following procedure to set up the pressures values to display in kPa.
1. On the CIC Setup screen, click CIC Defaults. The default pressures value is
mmHg.