Reload software
2026419-033E CIC Pro™ 11-7
Complete checkout procedures
Confirm images restored to correct drives
1. Log in as Administrator. See Log on to the CIC Pro center as Administrator on
page 4-3.
2. Right-click My Computer icon and select Manage.
3. Navigate to Storage > Disk Management.
4. Compare drive capacities as follows:
Disk 0 shows four partitions: C, D, E F. Disk 0 also has a smaller capacity
than Disk 1.
Disk 1 shows one partition: G and has a larger capacity than Disk 0.
NOTE
The smaller drive contains all the software and the larger drive contains only
full disclosure data.
Complete FRU checkout procedures
Refer to the following table for the applicable checkout procedure.
If this FRU is replaced Complete the applicable checkout procedures
Hard drive See page 9-17.
Flash drive (solid-state
drive)
See page 6-91.
CPU PCB (mother board) See page 9-14.