6-22 CIC Pro™ 2026419-033E
Configuration
4. At the command prompt, type stop and press Enter.
5. Select Start > Shut Down > Log off cic.
6. Click OK and immediately hold the Shift key down until a logon screen appears.
7. In the User name field, type administrator.
8. In the Password field, type admin1,3,5,7.
9. Select OK.
10. Double-click the My Computer icon on the desktop.
11. Navigate to D:\Documents and Settings\CIC.
12. Double-click the Favorites icon.
13. Right-click the browser favorite you want to delete and click Delete.
Configure set flags settings
This section describes the procedures required to configure the set flags settings:
Configure the NO COMM alarm setting, if applicable on page 6-22.
Configure force age setting, if applicable on page 6-23.
Configure multiviewer alarm audio setting, if applicable on page 6-24.
Configure ADU alarm audio setting, if applicable on page 6-26.
CAUTION
OUT-OF-UNIT ALARMS— If the CIC Pro center is configured to
alarm for out-of-unit alarms, any patient displayed on that out-of-
unit CIC Pro center, and is alarming, can have their alarms silenced
from the out-of-unit CIC Pro center.
NOTE
All clinical setting values must be received from the GE Clinical Application
Specialist (CAS) or from the nursing director of the care unit. Consult with your
CAS or hospital staff about the use of these functions and if they should be used
with your configuration.
Configure the NO COMM alarm setting, if applicable
NOTE
For NO COMM alarm behaviors, see the CIC Pro Clinical Information Center
Operator’s Manual.