Electromagnetic compatibility
2026419-033E CIC Pro™ A-3
Electromagnetic immunity
The CIC Pro center is intended for use in the electromagnetic environment specified
below. It is the responsibility of the customer or user to assure that the CIC Pro center
is used in such an environment.
NOTE
U
t
is the AC mains voltage prior to application of the test level.
RF Emissions (conductive)
EN 55011
Group 1
Class A
CAUTION
The equipment is suitable for use in all establishments other than domestic and
those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic Emissions
IEC 61000-3-2
Complies
Voltage Fluctuations/
Flicker Emissions
IEC 61000-3-3
Complies
Emissions test Compliance Electromagnetic environment – guidance
Immunity test IEC 60601 test level Compliance level Electromagnetic environment – guidance
Electrostatic Discharge
(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30%.
Electrical Fast Transient/
Burst
IEC 61000-4-4
± 2 kV for power supply lines
± 1 kV for input/output lines
± 2 kV for power supply
lines
± 1 kV for input/output
lines
Mains power should be that of a typical
commercial or hospital environment.
Surge
IEC 61000-4-5
± 1 kV differential mode
± 2 kV common mode
± 1 kV differential mode
± 2 kV common mode
Mains power should be that of a typical
commercial or hospital environment.
Voltage dips, short
interruptions and voltage
variations on power
supply input lines
IEC 61000-4-11
<5% U
t
(>95% dip in U
t
) for 0.5
cycles
<40% U
t
(>60% dip in U
t
) for 5
cycles
<70% U
t
(>30% dip in U
t
) for 25
cycles
<5% U
t
(>95% dip in U
t
) for 5 s
<5% U
t
(>95% dip in U
t
)
for 0.5 cycles
<40% U
t
(>60% dip in U
t
)
for 5 cycles
<70% U
t
(>30% dip in U
t
)
for 25 cycles
<5% U
t
(>95% dip in U
t
)
for 5 s
Mains power should be that of a typical
commercial or hospital environment. If the
user of the equipment requires continued
operation during power mains interruptions,
it is recommended that the equipment be
powered from an uninterruptable power
supply or a battery.
Power Frequency (50/60
Hz) Magnetic Field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital
environment.