Checkout procedures
2026419-033E CIC Pro™ 7-5
Verify settings with Check Centrals command
After introducing the upgraded or new CIC Pro center onto a verified compatible IX/
MC network, it is recommended that you run the Check Centrals diagnostic from the
CIC Pro center via Webmin as a final checkout. See page Pre-configuration
instructions on page 6-2 for instructions on running the Check Centrals utility.
Check and configure the speaker volume
WARNING
ALARMS — Do NOT rely exclusively on the audible alarm system
for Bedside Monitoring. Adjustment of CIC Pro center alarm
volume to a low level or OFF during Bedside Monitoring may result
in inability to hear the alarm and a hazard to the patient. Remember
that the most reliable method of Bedside Monitoring combines close
personal surveillance with correct operation of monitoring
equipment.
After connecting the monitor to the central station and/or nurse-
alert, verify the function of the alarm system. Repeat this
verification periodically, including a check of all connected
speakers.
The functions of the alarm system for monitoring of the patient must
be verified at regular intervals. Check speaker volume periodically
to ensure audio alarm functionality.
Activation code summary sheet CIC Setup > Licensing window