7-6 CIC Pro™ 2026419-033E
Checkout procedures
Complete the following procedure to configure the speaker volume for the CIC Pro
center:
NOTE
All clinical setting values must be received from the GE Clinical Application
Specialist (CAS) or from the nursing director of the care unit. Consult with your
CAS or hospital staff about the use of the various alarms in your environment and
the configuration of the alarm audio settings.
1. From the multi-patient viewer, click Setup CIC.
2. Click the Service Password tab.
3. In the Password field, type mms_com and press Enter.
4. At the command prompt, type setaudio<space>-d and press Enter. The current
speaker volume values for this CIC Pro center display.
5. Compare the speaker values displayed on your CIC Pro center to the following
HDM/5 speaker values:
6. If the values do not match, complete the following steps:
a. At the command prompt, type setaudio<space>-x%<space># and press
Enter. Repeat for each incorrect volume/value.
(x% = 10% to 100% and # = audio value setting)
b. Type setaudio<space>-d and press Enter to verify the audio levels.
7. Restart the CIC Pro center to apply your changes:
a. At the command prompt, type stop and press Enter.
b. From the Windows taskbar, click Start > Shut Down > Restart and press
Enter.
8. Verify the new settings prior to monitoring patients.
Volume Value
10% 2749
20% 3705
30% 4631
40% 5600
50% 6300
60% 7000
70% 10000
80% 12000
90% 16000
100% 20598