Configuration
2026419-033E CIC Pro™ 6-7
NOTE
In the hierarchy of multiple compatible CIC Pro center hardware and
software versions that co-exist, it is important to assign a set of the highest
MC IP addresses to the highest software version. For example, if you are
installing CIC v5.1.x to an existing CARESCAPE Network comprising of
CIC v4.0.7 or later, you must allocate a set of the highest MC IP addresses to
all of the CIC Pro centers hardware running CIC v5.1.x. No lower version
CIC Pro center can have an IP address on the MC network that is greater
than the IP address of a higher version CIC Pro center.
10. Verify that there is only one Time Master on the network and it is compliant with
step 9 by running the following command from any CIC Pro center currently on
the network:
lw -s “TIME MASTER”(lw<space>-s<space>”TIME<space>MASTER”)
11. If you have not already connected the service PC to the CIC Pro center, set up the
service PC’s network properties on page 4-8.
12. Proceed to the Configuration process checklist on page 6-7.
Configuration process checklist
NOTE
After making configuration changes, ensure the CIC Pro center is functioning
properly by completing all relevant checkout procedures before returning it to
clinical use. See Checkout procedures on page 7-1.
NOTE
If the System Resource Indicator ( icon in the bottom tray) turns yellow or
red during or after configuration, a pop-up message will appear. Take the
necessary action. Refer to Environment Monitor messages on page 8-19.
Disconnect CARESCAPE Network IX and MC networks
Disconnect from the CARESCAPE Network IX and MC networks on page 6-10.
Install licenses
Activate software licenses on page 6-10.
Configure Webmin-related settings
NOTE
All Webmin modules are static in nature. Since the Webmin service interface
session times out after 15 minutes, you must always refresh the browser to load
the latest page.
Log on to the CIC Pro center as Administrator on page 4-3.