Configuration
2026419-033E CIC Pro™ 6-91
Complete configuration checkout procedures
NOTE
Close all files before starting the checkout procedures.
Before using the CIC Pro center to monitor patients, you must verify proper operation
of this device in the patient care and networking environments.Use the following
checklist to thoroughly test the system for proper function and operation.
Check date and time on the network after installation of this current device. See
Pre-configuration instructions on page 6-2.
Verify that the Remote Service agent is enabled. See Enable or disable the
Remote Service Agent on page 6-18.
Check the unit defaults on page 7-3.
Check the laser printer connectivity. See Check status of installed printers on
page 7-3.
Check the digital writer connectivity. See Check status of installed printers on
page 7-3.
Check operation of the secondary display on page 7-4 (if applicable).
Check the status of installed licenses on page 7-4.
Verify settings with Check Centrals command on page 7-5.
Check and configure the speaker volume on page 7-5
Check operation of audible alarm tones on page 7-7.
Check access to all other care units on page 7-7.
Check the operation of the Citrix application viewed at the CIC Pro center on
page 7-7 (Hospital responsibility).
Check the hospital intranet browser functionality on page 7-7 (Hospital
responsibility).
Check the status of locked beds on page 7-7.
Check the language settings on page 7-7 (applicable for all non-English
applications only).
Check the pressures unit-of-measure on page 7-8 (for Chinese applications only).
Check the custom groupings for patient data on page 7-8.
Check the shortcuts to favorite CIC Pro center views on page 7-8.
Check MultiKM (Multimouse) operation on page 7-8.
Check current system settings on page 7-8.
Check full disclosure license type for all admitted in-unit beds on page 7-9.
Check full disclosure report printing on page 7-9.
Save backup defaults file on the USB memory stick on page 7-10 (Optional).
Check operation of the Watchdog countdown function on page 7-18.