EasyManuals Logo

GE MP100 Series User Manual

GE MP100 Series
314 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #117 background imageLoading...
Page #117 background image
Configuration
2026419-033E CIC Pro™ 6-23
NOTE
If the NO COMM is set to DISABLE, a visual only indicator will appear in
the multi-patient viewer window at the CIC Pro center. The clinical user will
not receive an audible notification. DISABLE is the factory default setting.
If the NO COMM is set to ENABLE, the clinical user will receive a crisis
level audible notification in addition to a visual NO COMM indicator in the
multi-patient viewer window.
When enabled, the system shall provide audible indication for NO COMM
in no more than 188 seconds from a loss of network communication with a
monitor.
Complete the following procedure to activate the NO COMM alarm on the CIC Pro
center:
1. If you have not already logged onto Webmin, Log on to the Webmin service
interface on page 4-7.
2. Click Configuration > Set Flags.
NOTE
The Set Flags Module screen shows the current set flags settings, which can
be modified.
3. In the NO COMM field on the Set Flags Module screen, click Enable.
4. Click Submit.
5. Restart the application. See Perform safe restart of the CIC Pro center on page 8-
30.
6. Verify the new settings prior to monitoring patients.
Configure force age setting, if applicable
If the force age setting is enabled, a default age range will not be displayed in the Age
field, and the New Patient button will be disabled. If the force age setting is disabled,
when a patient admit is attempted and no age is selected, a message will prompt the
user to enter the patient age. This setting is usually applicable for pediatric care areas.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the GE MP100 Series and is the answer not in the manual?

GE MP100 Series Specifications

General IconGeneral
BrandGE
ModelMP100 Series
CategoryMedical Equipment
LanguageEnglish

Related product manuals