6-24 CIC Pro™ 2026419-033E
Configuration
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.
Complete the following procedure to configure the Force Age setting on the CIC Pro
center:
1. If you have not already logged onto Webmin, Log on to the Webmin service
interface on page 4-7.
2. Click Configuration > Set Flags.
NOTE
The Set Flags Module screen shows the current set flags settings, which can
be modified.
3. In the FORCE AGE field on the Set Flags Module screen, click Enable.
4. Click Submit.
5. Restart the application. See Perform safe restart of the CIC Pro center on page 8-
30.
6. Verify the new settings prior to monitoring patients.
Configure multiviewer alarm audio setting, if applicable
This setting is applicable for mirror or out of care unit CIC Pro centers. This setting
will affect in-unit audible alarming.
NOTE
All clinical setting values must be received from the GE CAS or from the nursing
director of the care unit. Consult with your CAS or hospital staff about the use of
these functions and if they should be used with your configuration.