6-46 CIC Pro™ 2026419-033E
Configuration
NOTE
For more information on setting Telemetry Alarm Control Defaults, refer to the
telemetry system’s operator manual.
Complete the following procedure to configure the Telemetry Alarm Control
Defaults settings:
1. From the multi-patient viewer, click Setup CIC.
2. Log on to Setup CIC with service access on page 4-5.
3. From the CIC Setup screen, click Telemetry Alarm Controls Defaults.
4. On the Telemetry Alarm Control Defaults tab, change the settings as needed:
Option Function
Parameter Limits and
Alarm Levels
Set default parameter alarm limits and alarm levels for the following parameters: HR, NBP-S, NBP-D, NBP-M,
SPO2, SPO2-R, ST-I, ST-II, ST-III, ST-V, ST-V2, ST-V3, ST-V4, ST-V5, ST-V6, ST-aVR, ST-aVL, ST-aVF,
PVC, RR, and RR-Apnea.
Arrhythmia Alarm Levels Set default alarm levels for the following arrhythmia calls: ASYSTOLE, VFIB/VTAC, V TACH, V BRADY,
VT>2, ACC VENT, PAUSE, TACHY, BRADY, R ON T, COUPLET, BIGEMINY, TRIGEMINY, PVC,
IRREGULAR and ATRIAL FIB.
System Alarm Levels Set default alarm levels for the following system alarms: CHANGE BATTERY, OFF NETWORK, ARR
SUSPEND, LEADS FAIL, and PROBE OFF.