GE COMPANY
DIRECTION 5472001-1EN, REVISION 6OPTIMA CT680 SERIES AND OPTIMA CT670 INSTALLATION MANUAL
Page 176 Section 1.0 - Gantry Cover Removal
7.) Remove the Mylar (scan) window.
8.) Remove front cover.
a.) Disengage upper cantrell bracket on right side of the cover.
Figure A-14 Releasing Cover Brackets
A.) Disengage the locking mechanism on the upper cantrell brackets by using your
thumb to slide the trigger (red lever) back. This will release the locking mechanism
and allow the cantrell to be rotated upwards with steady and firm pressure.
B.) Disengage the rubber retaining straps on right side. See Figure A-15. You may find it
helpful to lift "up" on the cover to align the stud while attaching the rubber retaining
straps.
Figure A-15 Rubber Retaining Straps and Cover Locking Mechanism