GE Analytical Instruments ©2009 112 of 236 DLM 90488-01 Rev. A
Figure 18: The Maintenance Tab
Displaying Consumables Status
The Main screen displays an overview graphic that indicates the relative life status of the
Analyzer’s consumables: UV lamp, pump tubing, resin bed, acid, and oxidizer. The reagents’ status is
based on remaining calendar life and an estimate of remaining volume (based on the history of
usage), whichever is most limiting. Status indicators on the Main screen change color as the need
for replacement nears.
To display more detailed information about consumables status, follow these steps:
1. Select the Maintenance tab.
2. Press the Consumables button.
The status of each of the consumables is displayed, with an estimate of remaining life and either
the expiration date, the total days of usage, or the usage remaining.
If you need to install new consumables, follow the instructions in Chapter 7, "Maintenance" for each
consumable you install; also be sure to follow the instructions in the section called “Setting the
Installation or Expiration Date for New Consumables” on page 190.