GE Analytical Instruments ©2009 130 of 236 DLM 90488-01 Rev. A
• Press the Min Length button to set the minimum character length for a Password. The
default value is 3 characters.
• Press the Login Timeout button to set the number of minutes of allowable keypad
inactivity before users are required to re-enter their login information. The default value
is 5 minutes.
Changing User Passwords
All users can change their own Passwords by following these steps:
1. Log in with a valid User ID and Password.
2. Select the DataGuard tab.
3. Press the Change Password button.
4. Enter the old Password and press Enter.
5. Specify a new Password and press Enter.
6. Verify the new Password and press Enter.
Administrators can change the Password associated with any User ID by pressing the Password
button on the Edit User screen (“Editing User Information” on page 128).
Dealing with Forgotten Passwords
If users forget their Passwords, the administrator must assign new Passwords by following these
steps:
1. Log in to the Analyzer with an administrator-level User ID and Password.
2. Select the DataGuard tab.
3. Press the Edit User button.
4. Select the appropriate User ID from the list using the scroll buttons and press OK.
5. Press the Password button.
6. Specify a new Password and press Enter.
Note: Require a new user to change the Password upon log in so that it will be easy to
remember and unknown to the administrator.