Remove Internal Packing Materials
Some 900 On-Line Analyzers are shipped with packing materials inside the Analyzer case and can
be identified by a sticker that says “Remove” before use. If any packing materials are present,
remove them now.
Step 5: Fill the DI Water Reservoir
Fill the DI water reservoir with DI water, using the squeeze bottle provided in the accessories kit.
1. Fill the water bottle with DI water.
2. Remove the rubber inlet cover on the DI water reservoir.
3. Slide the bottle nozzle into the inlet hole, and gently squeeze the bottle. Be sure to
fill the reservoir until the water reaches the fill line, just below the inlet.
4. Replace the rubber inlet cover. Visually check for leaks from the reservoir after the
reservoir has been filled.
Figure 5: Filling the DI Water Reservoir
Note:
The DI water pump will not be properly primed unless you follow the instructions in
the section called “Powering the Analyzer On” later in this chapter.