GE Analytical Instruments ©2009 157 of 236 DLM 90488-01 Rev. A
13. If you have an ICR unit, return the lever to the Inline position.
Figure 29: The Autoreagent Verification Summary Screen
Reviewing Calibration and Verification History
If you need to review the results summary of past calibrations, follow these steps:
1. Press the Menu button and select the Maintenance tab.
2. Press the Calibrate button.
3. Press the Cal/Ver History button.
Press the Print History button to output the calibration history to the printer.
Press the Export History button to output the calibration history to the USB port or the
serial port in text (.csv) format.
If you want to review specific measurements from calibration or verification, you can display the
data history as you normally would. The Mode column lists TC Cal and IC Cal for calibration results,
TC Ver and IC Ver for verification results, and Sys Suit for system suitability results. These results
also are included when you export the data history.