GE Analytical Instruments ©2009 21 of 236 DLM 90488-01 Rev. A
Warnings
English
Warning
This symbol on the instrument indicates that the user should refer to
the manual for operating instructions
Warning
The IOS System and vial port contain two sharp needles designed to
pierce the septa of sample vials. Do not put fingers or inappropriate
materials into the IOS System or vial port.
Warning
In 900 Laboratory and Portable Analyzers, for continued protection
against fire hazard, replace fuse with same type and rating.
Warning
In 900 On-Line and Portable Analyzers, water in the IOS System may
be hot. Before inserting a vial into the IOS System, slide the door open
and wait 30 seconds to allow sample to completely drain. Inserting a
vial before draining can result in potentially hot water spray projecting
upward out of the IOS System.