Turbo Mode Specifications
When operated in Turbo mode, the Analyzer has the following specifications:
Linear range 0.20 ppb – 10 ppm TOC
Precision 2% RSD
Accuracy ±3%
Analysis Time: 4 seconds
Sample Flow Rate: 1.1 mL/min; 0.5 mL/min when TOC > 5 ppm
Instrument Delay (90% recovery) 3.5 minutes
Enabling Turbo Mode
Activation of Turbo sampling mode may be purchased directly from GE Analytical Instruments.
1. Select the Setup tab.
2. Press the Enable Turbo button.
3. Enter the Activation Code. If necessary, press the Number button to toggle the keypad
from letters to numbers.
4. Press the Enter button.
5. After Turbo mode has been enabled, you must turn Turbo mode on by following the
instructions in the next section.
Turning Turbo Mode On and Off
To turn Turbo mode on or off, follow these steps:
1. Press the Menu button and select the Setup tab.
2. Press the Configure button.