EasyManua.ls Logo

GE Sievers 900 Series - Turbo Mode Specifications; Enabling Turbo Mode; Turning Turbo Mode on and off

GE Sievers 900 Series
236 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2009 220 of 236 DLM 90488-01 Rev. A
Turbo Mode Specifications
When operated in Turbo mode, the Analyzer has the following specifications:
Linear range 0.20 ppb – 10 ppm TOC
Precision 2% RSD
Accuracy ±3%
Analysis Time: 4 seconds
Sample Flow Rate: 1.1 mL/min; 0.5 mL/min when TOC > 5 ppm
Instrument Delay (90% recovery) 3.5 minutes
Enabling Turbo Mode
Activation of Turbo sampling mode may be purchased directly from GE Analytical Instruments.
1. Select the Setup tab.
2. Press the Enable Turbo button.
3. Enter the Activation Code. If necessary, press the Number button to toggle the keypad
from letters to numbers.
4. Press the Enter button.
5. After Turbo mode has been enabled, you must turn Turbo mode on by following the
instructions in the next section.
Turning Turbo Mode On and Off
To turn Turbo mode on or off, follow these steps:
1. Press the Menu button and select the Setup tab.
2. Press the Configure button.

Table of Contents

Related product manuals