GE Analytical Instruments ©2009 139 of 236 DLM 90488-01 Rev. A
Autoreagent calibration standards sets include the standards shown in Table 15.
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
Autoreagent verification standards sets include the standards shown in Table 16.
*Note: 1ppm = 1 mg C/L, 1ppb = 1 µg C/L
Preparing for Calibration
Export Current Constants
Prior to performing any calibration procedure, export the current constants, in the event they need
to be re-loaded or referred to in the future. To export the constants, follow these steps:
1. Select the Maintenance tab.
2. Press the Advanced button.
Table 15: Standards Required for Autoreagent Calibration
Reagent Water
25 ppm TOC (as KHP)
50 ppm TOC (as nicotinamide)
25 ppm IC (as Na
2
CO
3
)
Table 16: Standards Required for Autoreagent Verification
Reagent Water
25 ppm TOC (as sucrose)
50 ppm TOC (as sucrose)
50 ppm IC (as Na
2
CO
3
)