Chapter 2. System Description
System Specifications
*
Linear range 0.03 ppb – 50 ppm TOC
Precision <1% of RSD
Accuracy ±2% or ±0.5 ppb, whichever is greater
Analysis Time 4 minutes
Sample Flow Rate (nominal) Analysis mode: 0.5 mL/min
Fast Flush (between samples): 1.0 mL/min
Turbo mode (optional): 1.1 mL/min
Required Sample Line Flow Rate: 50 mL - 300 mL/min (for On-Line Mode)
Power requirements 100-240 ±10% VAC, 100 watts, 50/60 Hz
Fuses No user-replaceable fuses; built-in circuit
breaker
Sample Temperature 1 °C to 95 °C via iOS** (withstands short-term
steam exposure)
Ambient Temperature 10 °C to 40 °C
Sample Pressure (in IOS System) Up to 250 psig
Normal Operating Environment Intended for indoor use only
Maximum Relative Humidity Up to 95%, noncondensing