GE Analytical Instruments ©2009 82 of 236 DLM 90488-01 Rev. A
15. Press the Back button to return to the Menu screen.
Setting the Clock and Time Zone
Set the Analyzer clock to ensure that all TOC measurements reflect the correct date and time:
1. Select the Maintenance tab.
2. Press the Clock button.
3. Press each date component button, enter the appropriate value, and press the Enter
button.
4. Press the Menu button, press the Advanced button, and then press the Advanced Setup
button.
5. (Optional) Press the Time Zone button, and specify the time zone via the two buttons on
this screen. Note that the time zone information you enter here does not modify the time to
which you set the clock in Step 3 above, but is merely intended to identify the time zone in
exported or printed data.
Time Zone — Allows you to set a text descriptor for the time zone. Usually this is a
three-letter code, such as “EST” for Eastern Standard Time or “GMT” for Greenwich
Mean Time.
GMT Difference — Allows you to enter the from Greenwich Mean Time. Use the
number pad and the +/- button to specify the . For example, for Eastern Standard Time
you should enter -5.00.
Naming the Analyzer Location (Optional)
You can assign a name to the Analyzer that displays with printed and exported data. This feature is
particularly useful if you have multiple Analyzers at your facility and want to easily distinguish data
collected from each unit. To assign a name, follow these steps:
1. Select the Maintenance tab.
2. Press the Advanced button.
3. Press the Advanced Setup button.