GE Analytical Instruments ©2009 188 of 236 DLM 90488-01 Rev. A
Figure 36: Brass-Colored Resin Bed Bracket
Replacing the In-Line Filter Element
To prevent clogging in on-line configurations, a filter is installed on the sample inlet line. The lifetime
of the filter element depends on the level of particles in the water samples. If monitoring the TOC of
the feed water (prior to purification), the filter element will need to be replaced more often than if
monitoring the water after purification.
If the filter element clogs too frequently, contact GE Analytical Instruments to receive help in the
selection of larger-capacity filters. As the filter is used, the flow rate of water through the IOS
System will decrease and can even stop. A simple way to determine if the filter element needs to be