GE Analytical Instruments ©2009 133 of 236 DLM 90488-01 Rev. A
Logging Out
When users have finished their activities on the Analyzer, logging out prevents a second user from
accessing the system under the first user’s account. This practice helps preserve the integrity of
the audit trail. Logging out also minimizes the risk of unauthorized access by other personnel.
To log out, press the Logout button on any display screen. If the Logout button is not displayed on
the current screen, press the Back or Menu button until any of the main tabs (Setup, Data, I/O,
Maintenance, or DataGuard) display, and then press the Logout button.
Table 9: Audit Trail Output Format
Field #
Field
Contents
Format/Notes Example
1 User ID 11 alpha-numeric characters (columns 1-11) ADMIN 01
2Entry ID
Number
9 numeric characters (columns 13-21) 693090237
3 Date and
Time
dd MMM yyyy hh:mm:ss (columns 23-33) 17 OCT 2004
4 Action 20 alpha-numeric characters (columns 44-63) PASSWORD SET
5 Old Value Alpha-numeric characters (columns 65-end of
line)
QUALITY 01
6 New Value Alpha-numeric characters (columns 65-end of
line)
QUALITY 01