GE Analytical Instruments ©2009 61 of 236 DLM 90488-01 Rev. A
Rubber plug for IOS waste line (only needed for 900 On-Line or 900 Portable Analyzers)
5. Reagent package containing:
Acid reagent cartridge
Oxidizer reagent cartridge
6. Certificate of Calibration
7. DataGuard Activation key (optional)
Additional Installation Equipment
To install the Sievers 900 On-Line TOC Analyzer, the following equipment also is required:
Mounting hardware to support the Analyzer (see “Step 3: Select a Location for the
Analyzer” on page 62)
Insulated wire (22-12 American Wire Gauge, rated to 300 Volts) for analog and alarm
outputs (optional; see “Installing the Analog Outputs and Alarms” on page 68)
Analog (4-20 mA) recorder (optional)
Insulated wire (18-12 American Wire Gauge, rated to 300 Volts) for AC power conduit (see
“Connecting to a Power Supply” on page 75)
An external circuit breaker or switch that disconnects both poles of the supply voltage,
rated appropriately (optional)
Grounding strap for ESD protection
Step 2: Complete the Identification Records
Complete the Identification Records section on “Identification Records” on page 2 by recording the
date of installation and the Analyzer serial number (found on the left side of the Analyzer).
Note: Do not start the Analyzer until the reagents have been installed and the DI water
reservoir has been filled with DI water.