GE Analytical Instruments ©2009 224 of 236 DLM 90488-01 Rev. A
Turbo Mode Maintenance Considerations
All regular maintenance functions described in Chapter 7, "Maintenance," apply to Analyzers
working in Turbo mode. In addition, if the Analyzer operates continuously in Turbo mode, the
operating life of the sample pump is reduced by half, to 2.5 years.
Table 19: Replacement Schedule for Consumables in Turbo Analyzers
Description Typical Operating Life
DI Water Reservoir Refill as needed (check when replacing reagents or UV lamp)
Oxidizer 3 calendar months maximum (depends on flow rate)
Acid 12 calendar months maximum (depends on flow rate)
UV Lamp 6 months
Sample Pump Tubing 12 months
Resin Bed Typically 12 Months (depends on water quality)
Sample Pump 2.5 years