GE Analytical Instruments ©2009 193 of 236 DLM 90488-01 Rev. A
Note: interval suggestions based on typical use. Reagent consumption depends on flow rate settings.
Note: the DI water reservoir should be checked and filled as necessary as part of every maintenance.
A copy of this worksheet should be completed for each year the Analyzer is in service. The worksheet is available from the GE
Analytical Instruments web site (www.GEInstruments.com).
Analyzer Serial number ______________ Date placed into operation __________
Dates of operation encompassed by this worksheet: from __________to _________
Table 18: Sievers 900 On-Line TOC Analyzer 1-Year Maintenance Worksheet
Item Period
Installation
Date
Expiration Date Notes/volume
Oxidizer supply Initial Install
Acid supply Initial Install
DI water reservoir Initial Install N/A
Oxidizer supply 3 months
Acid supply 3 months
DI water reservoir Initial Install N/A
UV Lamp 6 months
Oxidizer supply 6 months
Acid supply 6 months
DI water reservoir Initial Install N/A
Oxidizer supply 9 months
Acid supply 9 months
DI water reservoir Initial Install N/A
Pump Tubing 12 months
UV Lamp 12 months
Oxidizer supply 12 months
Acid supply 12 months
Resin Bed 12 months
DI water reservoir Initial Install N/A