GE Analytical Instruments ©2009 158 of 236 DLM 90488-01 Rev. A
Sample Flow Rate Calibration Worksheet
Company Name___________________________ Date ________________________
Analyst Name ____________________________ Firmware Version _____________
Analyzer Serial Number ____________________
Sample pump tubing replaced before sample flow rate calibration (Y/N): ______
(Line 1)
Enter data for the sample flow rate calibration:
Expected Measured % Diff Adjusted
mL/min: 0.50
________ ________ ________
Flow rate calibration pass/fail based upon the following criteria:
•The Adjusted sample flow rate is 0.48 to 0.52 mL/min
Sample Flow Rate Calibration Results:
Passed Failed
Sample Flow Rate Calibration Action:
Applied Canceled
% Diff =
Measured Flow Rate - 0.50
0.50
-----------------------------------------------------------------------
100%×