EasyManua.ls Logo

GE Sievers 900 Series - Page 121

GE Sievers 900 Series
236 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
GE Analytical Instruments ©2009 121 of 236 DLM 90488-01 Rev. A
Always stop analysis before opening the IOS System door to avoid generating erroneous
measurement data. If the flow sensor is On, opening the IOS System door while analysis is taking
place causes the flow sensor to activate and halt analysis. If the flow sensor is Off, opening the IOS
System door while analysis is taking place results in air being drawn into the Analyzer.
Warning
The IOS System contains two sharp needles designed to pierce the
septa of sample vials. Do not put fingers or inappropriate materials
into the IOS System.
Warning
Water in the IOS System may be hot. Before inserting a vial into the
IOS System, slide the door open and wait 30 seconds to allow sample
to completely drain. Inserting a vial before draining can result in
potentially hot water spray projecting upward out of the IOS System.

Table of Contents

Related product manuals