GE Analytical Instruments ©2009 160 of 236 DLM 90488-01 Rev. A
Single-Point Calibration Worksheet
Company Name___________________________ Date ________________________
Analyst Name ____________________________ Firmware Version _____________
Analyzer Serial Number ____________________
TOC Autozero Performed (Y/N): ______
(Line 1)
Reagent Flush Performed (Y/N): ______ (Line 2)
Concentration of Standard: ______ (Line 3)
Enter data for the TC Calibration:
Expected Measured % Diff Adjusted
Blank N/A
________ ________ ________
________ ________ ________ ________ ________
Enter data for the IC Calibration:
Expected Measured % Diff Adjusted
________ ________ ________ ________ ________
Calibration pass/fail based upon the following criterion:
• The % Diff value for each IC standard is ±10% or less.
Calibration Results:
Passed Failed
Calibration Action:
Applied Canceled
% Diff =
Measured Concentration - Expected Standard Concentration
Expected Standard Concentration
-----------------------------------------------------------------------------------------------------------------------------------------------------------------
100%×