GE Analytical Instruments ©2009 198 of 236 DLM 90488-01 Rev. A
13 - Acid Warning
Message: “The estimated acid life is less than 15 days. Order a new acid container.”
Explanation: Based on the expiration date, the Analyzer estimates that only 15 days of acid life
remain. Order a new acid container from GE now to avoid down time when the acid expires.
14 - Oxid Warning
Message: “The estimated oxidizer life is less than 15 days. Order a new oxidizer container.”
Explanation: Based on the expiration date, the Analyzer estimates that only 15 days of oxidizer life
remain. Order a new oxidizer container from GE now to avoid down time when the oxidizer expires.
15 - ID Board Warning
Message: “A problem has been detected with the ID Board. Please contact GE.”
Explanation: A communication error with the ID board has occurred. Turn the Analyzer off and on
and restart analysis. If the warning persists, contact Technical Support.
16 - Acid Warning
Message: “The acid reagent needs to be replaced.
Explanation: Based on the flow rate and history of usage, the Analyzer estimates that the acid
reagent has been completely consumed. Perform a visual check to confirm that there is no
remaining acid before continuing analysis and order a new acid container from GE now.
17 - Oxid Warning
Message: “The oxidizer reagent needs to be replaced.”
Explanation: Based on the flow rate and history of usage, the Analyzer estimates that the oxidizer
reagent has been completely consumed. If you use oxidizer, perform a visual check to confirm that
there is no remaining oxidizer before continuing analysis and order a new oxidizer container from
GE now.
18 - UV Lamp Warning
Message: “The UV lamp needs to be replaced.
Explanation: Based on the history of usage, the Analyzer estimates that the UV lamp has expired.
Order a UV lamp now. Continuing to use the expired UV lamp can result in incomplete oxidation and
poor measurements.
19 - Acid Warning
Message: “The acid reagent has exceeded its recommended life. Replace the container.”
Explanation: Based on the expiration date, the Analyzer estimates that the acid has expired. Order a
new acid container from GE now.”
20 - Oxid Warning
Message: “The oxidizer reagent has exceeded its recommended life. Replace the container.”
Explanation: Based on the expiration date, the Analyzer estimates that the oxidizer has expired. If
you use oxidizer, order a new oxidizer container from GE now.”