Load the back up in the "Flash Prom" to the PACS
Use ESD equipment to avoid electrostatic damage.
Electrical components
1. Turn off the power to the cabinet;
2. Remove the old “Flash Prom” with a prom lifter.
3. Press the “Cold Start” button next to the battery on the CPU. Hold the button during power up
of the cabinet, and release it when the red LED turns off.
4. Turn off the power to the cabinet; Install the new FlashProm.
1. Flash Prom
2. Light Emitting Diode (LED)
3. "Cold start" button
5. Turn on the power. Now the application program is downloaded from the FlashProm to the
CPU.
6. Start the machine and make sure that everything is correct.
Changing of operating language
The language on the control panel can be changed in two ways: via the control panel or CS1000. What
language can be used depends on which System PROM the PACS 3500 has.
It is possible to change between these languages: English, French, Spanish, German, Portuguese, Italian,
Dutch, Swedish, Finnish, Danish, Norwegian, Icelandic, Estonian, Latvian, Lithuanian, Polish, Czech,
Hungarian, Slovenian, Slovakian, Russian, Japanese, Romanian.
Page 125 of 166
<Doc_TEC><Doc_6001341502><Rel.A><Lang_en>

Table of Contents

Other manuals for Getinge 9125

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Getinge 9125 and is the answer not in the manual?

Getinge 9125 Specifications

General IconGeneral
BrandGetinge
Model9125
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals